Товаров
  
корзина пуста
  • Каталог оборудования Siemens

MultiRanger 200 HMI

  • Заказные данные
    (1)
  • Аксессуары
    (32)
  • Информационные материалы

    Информационные материалы

    MultiRanger 200 HMI is a versatile short to medium-range ultrasonic single and multi-vessel level monitor/controller for virtually any application in a wide range of industries.

    Область применения

    MultiRanger 200 HMI can be used with various materials, including, water, municipal waste, acids, woodchips, or on materials with high angles of repose. MultiRanger 200 HMI offers true dual point monitoring, digital communications with built-in Modbus RTU via RS 485, as well as compatibility with SIMATIC PDM, allowing PC configuration and set-up. MultiRanger 200 HMI features Sonic Intelligence advanced echo-processing software for increased reading reliability.

    MultiRanger 200 HMI will monitor open channel flow and features more advanced relay alarming and pump control functions as well as volume conversion.

    It is compatible with chemical-resistant EchoMax transducers that are approved for hostile environments.

    • Key Applications: wet wells, flumes/weirs, bar screen control, hoppers, chemical storage, liquid storage, crusher bins, dry solids storage

    Дизайн

    The MultiRanger 200 HMI is available in wall or panel mounting options.

    Чертеж

    MultiRanger 200 HMI, dimensions in mm (inch)

    Особенности

    • Easy to use HMI display with local four-button programming, menu-driven parameters, and Wizard support for key applications
    • English, German, French, Spanish, Chinese, Italian, Portuguese, and Russian texts on the HMI
    • Removable terminal blocks for ease of wiring
    • Digital input for back-up level override from point level device
    • Communication using built-in Modbus RTU via RS 485 and SIMATIC PDM configuration software
    • Compatible with SmartLinx system: PROFIBUS DP (cyclic access of process values only) and DeviceNET
    • Single or dual point level monitoring
    • Auto False-Echo Suppression for fixed obstruction avoidance
    • Differential amplifier transceiver for common mode noise reduction and improved signal-to-noise ratio
    • Level, volume, and flow measurements in open channels, differential control, extended pump control, and alarm functions
    • Wall and panel mounting options

    Схема подключения

    MultiRanger 200 HMI connections

    Технические данные

    Mode of Operation

     

    Measuring principle

    Ultrasonic level measurement

    Measuring range

    0.3 ... 15 m (1 ... 50 ft)

    Measuring points

    1 or 2

    Input

     

    Analog

    0 ... 20 mA or 4 ... 20 mA, from alternate device, scalable

    Discrete

    10 ... 50 V DC switching level Logical 0 ≤ 0.5 V DC Logical 1 = 10 ... 50 V DC max. 3 mA

    Output

     

    EchoMax transducer

    44 kHz

    Ultrasonic transducer

    Compatible transducers: ST-H and EchoMax series XPS-10, XPS-15/15F, and XRS-5

    Relays

    Rating 5 A at 250 V AC, non-inductive

    mA output

    0 ... 20 mA or 4 ... 20 mA

    • Max. load

    750 Ω, isolated

    • Resolution

    0.1 % of range

    Accuracy

     

    Error in measurement

    0.25 % of range or 6 mm (0.24 inch), whichever is greater

    Resolution

    0.1 % of measuring range1)or 2 mm (0.08 inch), whichever is greater

    Temperature compensation

    • -50 ... +150 °C (-58 ... +302 °F)
    • Integral temperature sensor
    • External TS-3 temperature sensor (optional)
    • Programmable fixed temperature values

    Rated operating conditions

     

    Installation conditions

     
    • Location

    Indoor/outdoor

    • Installation category

    II

    • Pollution degree

    4

    Ambient conditions

     
    • Ambient temperature (housing)

    -20 ... +50 °C (-4 ... +122 °F)

    Design

     

    Weight

     
    • Wall mount

    1.22 kg (2.68 lb)

    • Panel mount

    1.35 kg (2.97 lb)

    Material (enclosure)

    Polycarbonate

    Degree of protection (enclosure)

     
    • Wall mount

    IP65/Type 4X/NEMA 4X

    • Panel mount

    IP54/Type 3/NEMA 3

    Electrical connection

     
    • Transducer and mA output signal

    2-core copper conductor, twisted, shielded, 0.5 ... 0.75 mm² (22 ... 18 AWG), Belden 8760 or equivalent is acceptable

    • Max. separation between transducer and transceiver

    365 m (1 200 ft)

    Displays and controls

    60 x 40 mm (2.36 x 1.57 inch) LCD 240 x 160 pixels resolution

    Power supply

     

    AC version

    100 ... 230 V AC ± 15 %, 50/60 Hz, 36 VA (17 W)

    DC version

    12 ... 30 V DC (20 W)

    Certificates and approvals

    • CE, RCM2)
    • FM, CSAUS/C, UL
    • CSA Class I, Div. 2, Groups A, B, C, and D, Class II, Div. 2, Groups F and G, Class III (wall mount only)

    Communication

    • RS 232 with Modbus RTU or ASCII via RJ-11 connector
    • RS 485 with Modbus RTU or ASCII via terminal strips
    • Optional: SmartLinx cards for
      • PROFIBUS DPV1 (cyclic access of process values only)
      • DeviceNet


    1) Program range is defined as the empty distance to the face of the transducer plus any range extension

    2) EMC performance available on request

    Полезное